CPC 100
V 3.20
Preface - 3
►
All AC and DC output sockets of the
CPC 100
can carry life-hazardous voltage potential and
provide life-hazardous currents. Therefore:
►
While connecting cables to the
CPC 100
high-voltage or current outputs, or other
conducting parts that are not protected against accidental contact, press the Emergency
Stop button, and keep it pressed as long as an output signal is not absolutely necessary
for the test.
►
When connecting to the front panel input/output sockets, use wires with either 4 mm
safety “banana” connectors and plastic housing or, where applicable, with the especially
manufactured counterpart supplied by OMICRON electronics (e.g., for the
V2 AC
measuring input).
►
For the high-voltage and current output connectors on the left-hand side of the test set
(2kV AC, 400A DC and 800A AC, Ext. Booster), only use the specially manufactured
cables supplied by OMICRON electronics (refer to the chapter “Accessories” of the
CPC 100 Reference Manual available in PDF format on the
CPC 100 Toolset
CD-ROM
or the
CPC 100
Start Page
).
►
One end of the high-voltage cable has a coaxial safety plug that is certified for a voltage
level of 2kV AC. The other end is equipped with a safety banana plug that is insulated
with a shrink tube. When the
CPC 100
is switched on, consider this part of the cable a
hazard of electric shock.
►
If you do not use the high-current outputs
400A DC
or
800A AC
, or the high-voltage
output
2kV AC
, disconnect any cable that may be plugged in to these sockets.
Note:
The 400A DC or 800A AC outputs are not switched off by internal relays. Therefore,
if a test mode is selected that does not use either one of these two outputs, they still
generate current.
►
Do not stand right next to or directly underneath a connection point because the clamps may
fall off and touch you.
►
The red warning light on the
CPC 100
front panel indicates hazardous voltage and/or
current levels at the
CPC 100
outputs (red light “I” on or flashing). The green warning
light indicates that the
CPC 100
outputs are not activated.
Note:
If none or both warning lights are on, the unit is defective. Do not use it anymore.
►
Both of the high-current output sockets on the left-hand side of the test set (
400A DC
and
800A AC
) usually carry a relatively low-voltage potential. However, in case of an
internal insulation fault these outputs may carry up to 300 V. Consider these outputs life-
hazardous.
►
Always lock connectors properly.
►
The high-current cables for both the
800A AC
and
400A DC
outputs are equipped with
connection clamps at one end. If these connection clamps are attached to a test object’s
terminal that is situated above your head, make sure the clamp is securely attached. Due to
the weight of the cables the clamp may become loose and fall down.
►
Do not operate the
CPC 100
under ambient conditions that exceed the temperature and
humidity limits listed in ”General” on page Technical Data-3.
►
Do not operate the
CPC 100
in the presence of explosives, gas or vapors.
►
If the
CPC 100
or any add-on device or accessory does not seem to function properly, do
not use it anymore. If in doubt, call the OMICRON electronics hotline (refer to cover page of
this User Manual.
Power Supply
►
Supply the
CPC 100
only from a power outlet that has protective earth (PE).
►
An error message (313) appears if either the PE connection is defective or the power supply
has no galvanic connection to ground. In this case, make sure that the PE connection is
intact. If the PE connection is intact and the error message still appears, select the
Disable
ground check
check box at the
Device Setup
tab in the
Options
view.
►
Ground the isolating transformer outputs or generators used to supply the
CPC 100
on the
N (neutral) output or select the
Disable ground check
check box as described above.
►
Instead of supplying the
CPC 100
from phase - neutral (L1-N, A-N), it may also be supplied
from phase - phase (e.g., L1-L2; A-B). However, the voltage must not exceed 240V AC.
►
Fuse-protect the power supply (16A slow-acting fuse).
►
Do not use an extension cable on a cable reel to prevent an overheating of the cord; run out
the extension cord.
►
The connector “Ext. Booster” is
always
galvanically connected to mains, regardless of
whether an external booster is selected on the software tab
Options | Device Setup,
the
green warning light (0) is on, the outputs are turned off or the Emergency Stop button is
pressed. Handle with extreme caution. Do not use any booster cables other than those
supplied by OMICRON electronics.
The counterpart of the high-current sockets are locking connectors.
To lock these connectors safely, insert them carefully until you feel a
“click” position. Now they are locked. Confirm this by trying to pull them
out. This should not be possible now.
To remove the locking connectors, unlock them by pushing them in
completely first, and then pull them out.
Safety Instructions for the
CPC 100
and its Accessories
Содержание CPC 100
Страница 13: ...CPC 100 V 3 20 Introduction 8 Introduction...
Страница 25: ...CPC 100 V 3 20 Current Transformer 10 Current Transformer...
Страница 29: ...CPC 100 V 3 20 Voltage Transformer 4 Voltage Transformer...
Страница 65: ...CP TD1 CP TD1 10 CP TD1...
Страница 75: ...CP SB1 CP SB1 4 CP SB1...