CP CU1
CP CU1 - 3
The
CPC 100
must be configured for the
CP CU1
. Follow the steps below to configure the
CPC 100
:
1. Press the
Options
view selector button to open the
Options
window.
2. In the
External booster
combo box, select
CU 1
.
The
CT
and
VT
settings are set according to the built-in current and voltage transformers
automatically.
3. Set the current range of the
CP CU1
using the current range switch (see page CP CU1-2)
to the value configured by the
CPC 100
software.
Note:
Current range settings on the test card and on the
CP CU1
front panel must be the same.
Safety Instructions
►
Do not connect the measurement setup to overhead lines if there is a possibility of a
thunderstorm over any part of the lines to be measured. A lightning discharge to the line
under test can cause injury or possibly death of the operating staff.
►
It is strongly recommended to take all parallel lines out of service before proceeding.
Connecting the measurement setup to overhead lines with a life parallel system brings about
high-voltage hazards.
►
While the grounding switch at the near end of the power line is open, the area around the
CP GB1
in the range of 5 m/15 ft and around the
CP CU1
in the range of 2 m/5 ft is a
dangerous zone due to high-voltage and mechanical hazards. Do not enter the dangerous
zone. Keep the grounding switch open for a time as short as possible.
►
If you see or hear anything uncommon in the test equipment, e.g. noise of electrical
discharge or lightening of surge arrestors, close the grounding switch before touching the
measurement setup.
CPC 100
V1 AC
I AC
EXT. BOOSTER
CP CU1
V1 AC
I AC
I OUT
V SENSE
BOOSTER
CP GB1
(optional)
Test object
Dangerous zone
DANGER
Death or severe injury caused by high voltage or current
►
Only set the current range switch on the
CP CU1
front panel when the
CPC 100
is
turned off and the test object is connected to ground with closed grounding switch
near the measurement setup.
Measurement Setup
Configuring the
CPC 100
Connecting the
CPC 100
and
CP CU1
to Power Lines