![Omicron CPC 100 Скачать руководство пользователя страница 12](http://html2.mh-extra.com/html/omicron/cpc-100/cpc-100_user-manual_3990074012.webp)
CPC 100
V 3.20
Introduction - 7
Service
During operation, the
CPC 100
creates a log file with a user-definable logging level.
System Info
Displays system information.
1st Goal: Always Loading Certain Test Cards on System Start-Up
2nd Goal: Loading a Certain Test Card with Always the Same Values
With the exception of
Quick
, pressing an accelerator key opens the corresponding
Insert a new
test card
dialog box and lets you select the test card of your choice. Pressing
Quick
opens the
Quick
test card directly.
We recommend to set the logging
level to
Warning
.
Fill out one or more test card(s) of your choice with the parameters you need.
Change to Test Procedure Overview.
Press
Save As Default
You have now set the default for the
CPC 100
start-up.
Fill out the test card of your choice with the parameters you want to assign to that
card.
Put the focus to the test card’s tab.
Press
Save As Default
You have now changed the default for this test card type.
The command
Restore Defaults
at the
Options
tab
Device Setup
resets all
user-specific settings made in the
CPC 100
software to factory-defined defaults.
This includes the test card defaults and the test procedure default.
Quick
test card
Select from current transformer (CT)
test cards
Select from voltage transformer (VT) test cards
Select from power transformer test cards
Select from resistance test cards
Select from other test applications
The Options Menu
Customizing Your Working Environment
Accelerator Keys