CP SB1
CP SB1 - 3
Specifications
Weight and Dimensions
Environmental Conditions
Select Auto-tap
operation mode
Characteristic
Range
AC Input / V1 AC Output
max. 300 Vrms
DC Input
max. 6 A DC
Transformer High and Low Voltage
connections
max. 300 Veff between all connectors and ground
Tap Changer
Two potential-free contacts, short circuit-protected;
0-240 V AC (only AC permitted);
Overvoltage Category II;
Resistance per contact = max. 4
I continuous = 0.9 A rms
Supply
Via serial interface from the
CPC 100
(+15 V)
Overvoltage protection to case with
surge arrestors
All connections to Tranformer High and Low voltage
side; AC/DC Input; V1 AC / V DC Output
Weight
Dimensions (W x H x D)
CP SB1
test set
3.5 kg (7.7 lbs)
357 x 235 x 111 mm
(14.1 x 9.2 x 4.4")
test set & case
28.5 kg (62.7 lbs)
700 x 450 x 500 mm
(27.6 x 17.7 x 19.7")
Operating temperature
–10 … +55 °C (+14 … 131 °F)
Storage and transportation
–20 … +70 °C (–4 … 158 °F)
Max. altitude
2000 m
Humidity
5 … 95 % relative humidity; no condensation
Tested according to IEC 68-2-78
Vibration
Tested according to IEC 60068-2-6;
frequency range 10 … 150 Hz; acceleration 2 g
continuous (20 m/s
²
); 20 cycles per axis
Shock
Tested according to IEC 60068-2-27 (operating mode);
15 g / 11 ms, half-sinusoid, 3 shocks in each axis
EMC
EN 61326-1 Class A
IEC 61326-1 Class A
FCC Subpart B of Part 15 Class A
EN 61326-1
IEC 61326-1
Safety
EN 61010-1
IEC 61010-1
UL 61010-1
Suitable for usage according to
IEEE 510, EN 50191 (VDE 0104), EN 50110-1
(VDE 0105 Part 100)
Housing and transport case
IP20 according to EN 60529
Measurement Setup
Technical Data
Содержание CPC 100
Страница 13: ...CPC 100 V 3 20 Introduction 8 Introduction...
Страница 25: ...CPC 100 V 3 20 Current Transformer 10 Current Transformer...
Страница 29: ...CPC 100 V 3 20 Voltage Transformer 4 Voltage Transformer...
Страница 65: ...CP TD1 CP TD1 10 CP TD1...
Страница 75: ...CP SB1 CP SB1 4 CP SB1...