RTD and Thermistor Specification, Output (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1
Range (° C)
Output/Input
RTD Type
Minimum
Maximum
90 days
1 year
Pt 385, 1000
Ω
-200 °C
-80 °C
0.06 °C
0.07 °C
-80 °C
0 °C
0.06 °C
0.08 °C
0 °C
100 °C
0.07 °C
0.08 °C
100 °C
260 °C
0.07 °C
0.08 °C
260 °C
300 °C
0.07 °C
0.09 °C
300 °C
400 °C
0.07 °C
0.09 °C
400 °C
600 °C
0.08 °C
0.09 °C
600 °C
630 °C
0.08 °C
0.09 °C
Ni 120, 120
Ω
-80 °C
0 °C
0.02 °C
0.02 °C
0 °C
100 °C
0.02 °C
0.02 °C
100 °C
260 °C
0.01 °C
0.02 °C
Cu 427, 10
Ω
-100 °C
260 °C
0.30 °C
0.38 °C
YSI 400
15 °C
50 °C
0.005 °C
0.007 °C
1. 2-wire output
94
Содержание CL3001