*TST?
Runs a series of self-tests and returns a "0" for pass or a
"1" for fail. If any faults are detected, they are logged into
the fault queue where they can be read by the FAULT?
query.
*WAI
Prevents further remote commands from being executed
until all previous remote commands have been completed.
External Connection Commands
Command
Description
FUNC?
Returns the present output, measurement, or calibration
function selected on the isolated and primary displays, in
that order.
HART?
Returns the HART resistor setting for the isolated
milliamp range, ON or OFF.
HART_OFF
Turns off the HART resistor on the isolated milliamp
range.
HART_ON
Turns on the HART resistor on the isolated milliamp
range.
ISO_PRES_UNIT
Sets the isolated display pressure units.
ISO_PRES_UNIT?
Returns the isolated display pressure units.
LOOP_POWER?
Returns the 24V loop power setting for the isolated
milliamp range, ON or OFF.
LOOP_POWER _OFF Turns off the 24V loop power on the isolated milliamp
range.
LOOP_POWER _ON
Turns on the 24V loop power on the isolated milliamp
range.
PRES_UNIT
Sets the primary display pressure units.
PRES_UNIT?
Returns the primary display pressure units.
RTD_TYPE
Sets the Resistance Temperature Detector (RTD) type.
RTD_TYPE?
Returns the Resistance Temperature Detector (RTD) type.
TC_REF
Selects the internal temperature sensor or an external
reference value for cold junction compensation of
thermocouple (TC) source and measurement.
TC_REF?
Returns the source of the temperature being used for
cold junction compensation of thermocouple (TC) source
and measurement.
TC_TYPE
Sets the thermocouple (TC) type.
TC_TYPE?
Returns the thermocouple (TC) type.
TSENS_TYPE
Sets temperature mode, RTD or TC.
TSENS_TYPE?
Returns the temperature mode.