a) Connect the pressure module to the CL3001 as shown in figure
19.
Figure 19 - Isolated Pressure Module Connection
b) Press the
key. The CL3001 automatically senses which
pressure module is attached and sets its range accordingly.
c) If necessary, press the
key again to cycle through the
pressure units until the desired one is displayed.
d) Before attaching the module to the pressure source, zero the
module as described in the instruction sheet that came with the
module. Procedures vary, but all end with pressing the
and
keys.
e) Attach the module to the pressure source according to the
instruction sheet that came with the module, taking care to follow
all safety precautions when dealing with high pressures.
6. Output Setpoints
Nine preset output setpoints may be stored and recalled for each of
the following output modes:
• Voltage
• Current
• each thermocouple type, including millivolts
• each RTD type, including each of the five custom curves.
They may be recalled on an individual basis, or as an automatic up
and down cycle with a configurable dwell time between each
setpoint. The automatic cycle feature always starts at setpoint
number 1, stepping up to a user specified ending setpoint number,
then back down in reverse order, and then repeats.
31