OPER
FAULT?
This example demonstrates setting the CL3001
output to 1.1 volts, waiting for the output to settle
before activating the output and checking if the
sequence of commands completed successfully.
ZERO_MEAS
IEEE-488
RS-232
Sequential
Overlapped
This command zeros a pressure module, the thermocouple mV
offset, or the RTD ohms offset. For detailed zeroing instructions see
section 4.3 for RTD ohms, section 4.7 for thermocouple millivolts, or
section 4.9 for pressure.
Parameter:
<None> for RTD ohms, thermocouple millivolts,
or a pressure module that does not measure
absolute pressure
<value> is barometric pressure for absolute
pressure modules, expressed in the currently
selected pressure units
Response:
<None>
Example:
ZERO_MEAS
If thermocouple millivolts are currently selected,
this example zeros the offset.
ZERO_MEAS?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the zero offset for pressure modules,
thermocouple millivolts, or RTD ohms.
Parameter:
<None>
Response: <zero
offset>,<units>
where < zero offset > is the current offset.
and where <units> is one of the following:
OHM
Ohms
V
DC volts (thermocouple millivolts)
or
one of the pressure units listed
with the PRES_UNIT? command
Example:
ZERO_MEAS?
1.060000E-01,PSI
This example indicates that the pressure zero for
the attached module is 0.106 psi.
X
X
X
X
X
X
85