
ISO_MEAS
IEEE-488
RS-232
Sequential
Overlapped
This command sets the isolated measurement type.
Parameter:
<value>
where <value> is one of the following:
DC10V
measure DC voltage, 10V
range
DC100V
measure DC voltage,
100V range
DCI
measure DC current
PRESSURE measure
pressure
Response:
<None>
Example:
ISO_MEAS DCI
This example sets the isolated measurement to
DC current.
ISO_PRES_UNIT
IEEE-488
RS-232
Sequential
Overlapped
This command sets the isolated pressure unit.
Parameter:
<value>
where <value> is one of the following:
PSI
pounds per square inch
INH2O4C
inches of water at 4 °C
INH2O20C
inches of water at 20 °C
INH2O60F
inches of water at 60 °F
CMH2O4C
centimeters of water at 4
°C
CMH2O20C
centimeters of water at 20
°C
MMH2O4C
millimeters of water at 4
°C
MMH2O20C
millimeters of water at 20
°C
BAR
bars
MBAR
millibars
KPA
kilopascals
MPA
megapascals
INHG
inches of mercury at 0 °C
MMHG
millimeters of mercury at
0 °C
KG/CM2
kilograms per square
centimeter
Response:
<None>
X
X
X
X
X
X
66
Содержание CL3001
Страница 103: ......