10.4
Remote Command Listing
The following is an alphabetical list of all CL3001 remote commands
and queries, including the common commands and the device-
dependent commands. Each command title includes a checkbox
that indicates the remote interface applicability, IEEE-488 and/or RS-
232, and the command group, Sequential or Overlapped; see
section 9.6.1 for a description of these terms.
*CLS
IEEE-488
RS-232
Sequential
Overlapped
Clear Status command. This command clears the ESR, the error
queue, and the RQS bit in the status byte. This command terminates
any pending operation complete commands, *OPC or *OPC?.
Parameter:
<None>
Response:
<None>
Example:
*CLS
This example clears the ESR, the error queue, and
the RQS bit in the status byte.
*ESE
IEEE-488
RS-232
Sequential
Overlapped
Event Status Enable command. This command loads a byte into
the Event Status Enable (ESE) register. See the Event Status Enable
Register (ESE) description in section 9.7.
Parameter:
<value>
where <value> is the decimal equivalent of the
ESE byte, 0 to 255
Response:
<None>
Example:
*ESE 140
This example loads decimal 140 (binary
10001100) to enable bits 7 (PON), 3 (DDE) and 2
(QYE).
*ESE?
IEEE-488
RS-232
Sequential
Overlapped
Event Status Enable query. This command returns the contents of
the Event Status Enable (ESE) register. See the Event Status Enable
Register (ESE) description in section 9.7.
Parameter:
<None>
Response:
<value>
where <value> is the decimal equivalent of the
X
X
X
X
X
X
X
X
X
62