To set a setpoint:
a)
Select the output mode.
b)
Enter the output value for the setpoint.
c)
Press the
and
keys to select the SET function.
d)
At the setpoint number selection prompt "SET POINT#",
press the numeric key, 1 to 9, corresponding to the
setpoint to be set.
e)
If the automatic cycle feature is to be used, care should be
taken to order the setpoint values in an appropriate
manner. It always cycles between setpoint number 1 and
a user specified ending setpoint number. The values in
the cycled group of setpoints should be entered with this
in mind. Any random setpoints used for individual checks
can then be located after the usual ending setpoint
number.
To recall a single setpoint:
a)
Select the output mode.
b)
Press the
and
keys to select the RECALL
function.
c)
At the setpoint number selection prompt "RECALL SPT#",
press the numeric key, 1 to 9, corresponding to the
setpoint to be recalled.
To start an automatic setpoint cycle:
a)
Select the output mode.
b)
Press the
and
keys to select the AUTOSET
function.
c)
At the ending setpoint number selection prompt "AUTO
SET POINT", press the numeric key, 1 to 9, corresponding
to the ending setpoint number for the cycle.
d)
At the dwell time prompt "DWELL TIME", "5-500?", enter
the number of seconds, 5 to 500, to dwell at each setpoint
value, followed by the
key.
e)
The
key can be used at any time during the cycle
without stopping it.
f)
Press any other key to terminate the cycle.