c) Press the
key to select the desired pressure units for
display.
d) Before attaching the module to the pressure source, zero the
module as described in the instruction sheet that came with the
module. Procedures vary, but all end with pressing the
and
keys.
e) Attach the module to the pressure source according to the
instruction sheet that came with the module, taking care to follow
all safety precautions when dealing with high pressures.
5. Isolated Inputs
5.1 Voltage Input
The CL3001 can measure DC voltages from 0 V to 100 V, using the
following two ranges for maximum accuracy: 10 V, and 100 V.
a) Disconnect any test leads from external devices.
b) Press the
key to select isolated DC voltage and current
input mode, if not already selected. If the desired DC voltage
mode is not displayed, press the key again to cycle to the
desired DC voltage mode.
c) Connect the unit under test to the isolated voltage/current input
terminals of the CL3001 as shown in figure 17.
Figure 17 - Isolated DC Voltage Input Connection
29