Managing the Datalogger and Data Communication Features
Part # 910-250C
77
8.7
Recalling Transducer Calibrations and Waveforms
The EPOCH 4PLUS datalogger is designed to let you quickly review the contents of saved
files and then recall saved transducer calibrations to the live screen.
Note:
The EPOCH 4PLUS has a total of nine dedicated transducer calibration
slots. These are located in files 1-9 of the datalogger and are indicated by an
asterisk (
*)
next to the file number. These can only store one calibration
(with complete setup parameters and waveform) in each file. They are
designed to store the most frequently used calibration setups and can be
recalled using only two key presses. See Saving and Recalling Quick Recall
Calibrations on page 8.
To recall a saved calibration (in Files 10 and greater), follow these steps:
1. Press
to enter the Memory screen.
2. Use the slewing keys to select the file where the desired calibration is located.
3. When the file is highlighted, press [F1] to enter the Contents screen.
4. Use the slewing keys to scroll through the selected file. Only IDs with saved
waveform/parameter setups can be recalled to the live screen. The ID name being
viewed displays at the top of the screen. As you scroll through the file, this ID will
continually change.
a. When an ID is listed that has a corresponding waveform and setup parameters saved
with it, a Recall selection is available by pressing [F2].
b. The Recall selection disappears when an ID is listed that contains a measurement
value only (depth, amplitude, or TOF).
Figure 41
Recalling a Selection
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...