34
EPOCH 4PLUS
4.4
Using the Pulser Key
The following pulser receiver and filter parameters can be adjusted on the EPOCH 4PLUS
keypad:
•
Waveform Rectification
•
Pulser Energy
•
Pulser Damping
•
Pulser Test Mode
•
Filter Selection
•
Pulser Type (square wave or spike excitation)
•
Pulser Frequency (tuned for square wave pulser only)
To access the available parameters, follow these steps:
1. Press [PULSER] multiple times to toggle through the several features.
2. Press [2nd F], [PULSER] (FILTER) to go directly to the filter parameter.
Each pulser-receiver and filter characteristic adjustments can be done in either full A-
Scan or Split Screen.
•
In the full A-Scan, the current parameter being adjusted appears at the top left of
the screen and directly above the waveform.
•
In the Split Screen mode, the parameter is highlighted on the lower half of the
EPOCH 4PLUS screen.
Note:
For reviewing the adjustments in this manual, the Split Screen mode was
used.
To view how the selections of the pulser receiver affect a waveform, couple a transducer
to a sample or test block, which produces an echo on the top half of the screen.
4.4.1
Waveform Rectification
The EPOCH 4PLUS can operate in one of four different rectification modes:
•
Full Wave
•
Half Wave Positive
•
Half Wave Negative
•
RF (unrectified)
The RF mode is not active while operating in DAC or TVG mode, or Peak Memory.
To select a waveform rectification mode, follow these steps:
1. Press [PULSER] once to display the current screen rectification. This is highlighted
and displayed at the bottom of the screen.
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...