128
EPOCH 4PLUS
9.10.4
Using Gate Alarms
Individual gate alarms can be set while Floating Gate is active. In general practice, select
the minimum depth alarms to monitor for thinning areas in the material.
9.11
Curved Surface Correction (CSC)
The Curved Surface Correction (CSC) option for the EPOCH 4PLUS provides accurate
sound path values when inspecting curved surfaces such as pipes. When you enter the pipe
diameter and thickness, soundpath information is adjusted accordingly when using an
angle beam probe to inspect the circumference of the curved surface.
1. Press [OPTION].
2. Use the slewing keys to highlight CSC from the menu and press [ENTER].
A
!
appears, indicating that the CSC option is active. A cursor appears over the outer
diameter value.
3. Use the slewing keys to enter the desired diameter.
4. Press [F1] and then [OPTION] to return to the live screen. A CSC prompt is displayed
to the right of the A-Scan whenever the EPOCH 4PLUS is in CSC mode.
5. Calibrate the EPOCH 4PLUS and be sure to enter the appropriate shear wave angle.
Correct soundpath information is displayed when an echo breaks a gate.
Note:
The CSC option is linked to the EPOCH 4PLUS memory. If a calibration or
waveform is saved in CSC mode, the instrument will be in CSC mode when
that data set is recalled. A “c” is displayed on the Memory screen after the ID
to signify the calibration or waveform was saved in CSC mode.
To exit the CSC Mode, follow these steps:
1. Press [OPTION] and highlight CSC.
2. Press [ENTER] to display the CSC setup menu.
3. Press [ENTER] to disable the option and deselect the check box.
4. Press [F1] to return to the Option menu.
9.12
Wave Analysis
The Wave Analysis software option is useful in many specialized applications where
operators need to select particular points along an A-Scan waveform and then obtain a
timing (thickness) measurement between these two points. In addition, the software
displays the dB difference between the two points (for example, measuring the scale/
oxide layer buildup on the inner diameter of metal pipes). For this application, it is
generally recommended to use Olympus NDT
TM
M116 or V116 transducers.
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...