Adjusting the Pulser Receiver
Part # 910-250C
39
4.4.7
Pulser Frequency Selection
Pulser Frequency Selection is applicable only when the square wave pulser is selected.
The frequency selection is designed to tune the square wave pulser to obtain the best
performance for the transducer being used.
The Pulser Frequency Selection is viewable on the display only when the EPOCH 4PLUS
is in the Square Wave Pulser mode. In Spike mode, the selection is disabled.
To select Pulser Frequency Selection, follow these steps:
1. Press [PULSER] multiple times until the FREQ selection is highlighted.
2. Use the function keys for direct access or the [ ],[ ],[
] and [
] keys to select the
desired frequency.
3. Choose the frequency that is closest to the frequency of the transducer being used.
Note:
Actual results may vary. It is recommended that you try various settings
when evaluating a test sample in order to maximize ultrasonic performance.