Adjusting the Pulser Receiver
Part # 910-250C
31
4
Adjusting the Pulser Receiver
This chapter describes how to adjust the pulser receiver.
Topics are as follows:
•
Adjusting System Sensitivity
•
Using AUTO-80%
•
Setting Gain Reference Level and Adding Scanning Gain
•
Using the Pulser Key
4.1
Adjusting System Sensitivity
To adjust the instrument sensitivity, follow these steps:
1. Press [GAIN].
2. Use the function keys to select preset values for course adjustment.
3. Move the gain upward or downward with the [ ] and [ ] keys for fine adjustments.
The current system sensitivity is displayed and highlighted in the upper left corner.
4. Adjust gain in 0.1 dB increments by pressing the [ ] or [ ] key.
5. Hold the [ ] or [ ] key to gradually increase the slewing rate. Total system sensitivity
is 110.0 dB.
4.2
Using AUTO-80%
AUTO-80% quickly adjusts the gain (dB) setting, bringing the peak of a reference echo to
80% full screen height. AUTO-80% is especially useful with general welding
applications and codes that require quick gain adjustments to bring an echo to 80% screen
height.
You can use AUTO-80% to bring an echo to 80% full screen height in either Gate 1 or
Gate 2.
To use Auto-80% with Gate 1, follow these steps:
1. Press [GATE 1].
2. Use the green slewing keys to position the gate over the desired echo. Be sure the echo
breaks the gate threshold.
3. Press [F5]. The peak of the highest amplitude echo breaking the Gate 1 threshold is
brought to 80% screen height.
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...