Calibrating the EPOCH 4PLUS
Part # 910-250C
63
4. Position Gate 1 so that the leading edge of the backwall echo from the known
thickness step is exceeding the gate threshold. A thickness reading appears in large
text above the A-Scan.
Figure 26
Calibrating for a Thin Block Using a Dual Element Transducer
5. Press [ZERO OFFSET] once the reading is steady. The screen freezes and a pop-up
box appears on the screen.
Use the alphanumeric keypad to enter the exact known thickness of the test sample.
For this example, press [1], [0], [0]. The entry appears in the pop-up box.
If you enter an incorrect number, press [0] multiple times to clear the entry and then
type the correct thickness.
Figure 27
Entering a Thin Standard Value
6. Press [CALIBRATION]. The display returns to the live A-Scan. Couple the
transducer to the THICK calibration block step. For this example, the transducer is
coupled to the 0.500" step.
7. Position Gate 1 so that the leading edge of the backwall echo from the known
thickness step is exceeding the gate threshold. A thickness reading appears in large
text above the A-Scan.
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...