Figure 96
Auto-Freeze
Option 1 – HOLD MODE
When this mode is selected, when the EPOCH 4PLUS Auto-Freezes, the display stays
frozen until the user presses the FREEZE key.
Option 2 – TIME MODE
When this mode is selected, when the EPOCH 4PLUS Auto-Freezes, the display stays
frozen for a user-defined period of time. When TIME MODE is activated, the user must
set the amount of time for the EPOCH 4PLUS to freeze using the up and down arrow keys.
You may set the EPOCH 4PLUS to freeze for durations between 1 and 30 seconds in one
second intervals.
Before the option becomes active, the user must press [2
nd
F], [FREEZE]. It is
recommended that you adjust the instrument parameters such as Gain and Pulser Settings
before activating the Auto-Freeze option. The Auto Freeze option can be deactivated at
any time by pressing [2
nd
F], [FREEZE] again.
In order to make the EPOCH 4PLUS automatically freeze, the user must setup the proper
gate alarm(s). The EPOCH 4PLUS will Auto-Freeze when a gate alarm is triggered. For
example, if the user positions Gate 1 above the first echo location for a Spotweld
inspection and places a positive logic alarm on Gate 1, when the echo breaks Gate 1 the
EPOCH 4PLUS will Auto-Freeze. This function will work for all alarm types on both
Gate 1 and Gate 2. Auto-Freeze is not active for Interface Gate alarms.
When the Auto-Freeze option has been setup and activated, a snowflake shaped icon
appears on the right side of the display where the standard FREEZE “F” icon would
appear, as shown below:
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...