Chapter 5 Operation
65
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
5.3 Automatic mist & smoke evacuation
system/function (when using the compatible high
flow insufflation unit)
CAUTION
When using the automatic mist & smoke
evacuation system/function, also read the
instruction manual for the compatible high flow
insufflation unit (UHI-2 or UHI-3).
If exhaust does not stop 5 seconds after
stopping the output, the compatible high flow
insufflation
unit
may
be
malfunctioning.
Disconnect the communication cable and
contact Olympus.
Confirm that the electrosurgical generator is
turned ON. Otherwise, the automatic mist &
smoke evacuation system/function of the
compatible high flow insufflation unit will not
work.
Press the power switch of the compatible high flow insufflation unit (see Fig.
5.1).
While the handswitch or footswitch pedal is pressed, the automatic mist &
smoke evacuation system/function of the UHI-2 or UHI-3 is activated
simultaneously as the output. The exhaust continues for 5 seconds even
after the output is stopped.
For the connection method, see chapter 3.5, “Automatic mist & smoke
evacuation system/function (when using the compatible high flow insufflation
unit)”.
Figure 5.1
High flow insufflation unit (UHI-2)
High flow insufflation unit (UHI-3)