10
Important Information – Please Read Before Use
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
plugs and sockets, refer to chapter 3.8,
“Connection of HF instruments”.
Before use, the electrosurgical generator and
accessories must be examined for damage. All
cables and its plugs must be free of scratches
and cracks. Cables and accessories with
damaged insulation or connections must not be
used.
Electric shock
WARNING
Grounding failure
To prevent the risk of electric shock, the housing of
the electrosurgical generator must be grounded.
Always connect the power cord plug to a properly
grounded wall outlet. Do not use a 3-pin / 2-pin
adapter, as it can impair safe operation of the
electrosurgical generator.
WARNING
User shock
To prevent user shock, malfunction and damage of
the electrosurgical generator, keep liquids away
from all electrical equipment. If liquid gets on or
into the electrosurgical generator, terminate
operation immediately and contact Olympus.
CAUTION
Injury during servicing
When the housing is opened, there is a danger of
electric shock. The electrosurgical generator must
only be serviced by authorized technicians.
Burns
WARNING
Patient and / or user
The maximum output voltage characteristics of
the electrosurgical generator are shown in the
diagrams in “Output characteristics” in the
Appendix. When setting the power level, first
set it to a low level and increase it gradually. If
the output is initially set to a high level, the
electrode’s insulation may be damaged and
cause user and / or patient burns. However,
certain modes may present an unacceptable
risk at low output power settings. For example,
with the PulseCut fast mode or PulseCut slow
mode, the risk of an excessive thermal effect
rises if the output power setting is too low.
Therefore, it is recommended that you perform
basic testing before using the electrosurgical