14
Important Information – Please Read Before Use
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
WARNING
Ignitable cleaning- and disinfection agents
Flammable agents used for cleaning and
disinfection must be allowed to evaporate
before the electrosurgical generator is used.
Also ensure that flammable solutions are
neither on the patient’s skin (e.g. under neutral
electrode) nor in the patient’s body cavity when
the electrosurgical generator is used.
Non-flammable agents should be used for
cleaning and disinfection wherever possible.
WARNING
Ignitable materials
If absorbent cotton or gauze is used during the
procedure, it can be ignited by a spark
generated in the normal operation of the
equipment.
When performing electrosurgery, sparks occur
which could lead to burning or deflagration of
combustible materials.
Body hair is flammable. Water soluble surgical
lubricating jelly may be used to cover hair close
to the surgical site to decrease flammability.
WARNING
Risk of fire
Disconnect the power plug before changing the
fuses! Replace fuses as marked. The fuses must
only be replaced by authorized technicians!
Procedural hazards and complications
DANGER
Procedural hazards and complications
The safety of electrosurgery will be greatly
enhanced by a thorough knowledge of the
medical literature on the subject. Study of
specific information on the hazards and
complications of the procedure in question is
especially recommended.
To respond to possible patient bleeding,
prepare at least one of the following three
haemostatic procedures: coagulation, clipping
or local injection.
To prepare for possible accidents, emergency
equipment for life-saving, intubation and
appropriate pharmaceuticals should be located
in or near the procedure room.
Always
prepare
a
spare
electrosurgical
generator or an alternative procedure to avoid