ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
Mode
Touch-
screen
Crest
factor* at
rated
load
Default
power
level
Min.**
power
level
[W]***
Max.**
power
level
[W]***
Default
effect
Num-
ber of
effects
Rated
load
[Ohm]
Max.**
output
voltage
[V
p
]
Waveform
BiSoft
Coag
N/A
55
1
120
2
3
75
220
Continuous
sinusoidal
alternating
current
Auto
Coag
N/A
50
1
120
2
3
75
220
Continuous
sinusoidal
alternating
current
Saline
Coag
N/A
120
10
200
2
3
75
230
Continuous
sinusoidal
alternating
current
Hard
Coag
N/A
50
1
120
2
3
25
220
Continuous
sinusoidal
alternating
current
RFCoag
N/A
30
1
50
w/o
RCAP
w/
RCAP
75
220
Continuous
sinusoidal
alternating
current
Bipolar
coagulation
Fine
Coag
N/A
40
1
40
1
1
50
220
Continuous
sinusoidal
alternating
current
*Crest factor at rated load specified for modulated sinusoidal alternating current modes only.
**min. = minimum; max. = maximum
***The power level is the maximum power (watts) which can be applied during the activation of the
electrosurgical generator. The actual applied power depends on the tissue characteristics (e.g.
resistance).
All cutting modes (except BlendCut) are supported by the High Power Cut
Support (HPCS) according to figure 9.1. The high power cut support optimizes
the start of the cutting procedure by applying high power to the tissue to support
immediate spark ignition and reduces the risk of mechanical cutting.
Figure 9.1
Example:
High Power Cut Support during PulseCut fast / slow
(t
1
is the pulse delay time)
P
H
P
C
S
C
U
T
C O A G
t
1
Spark
Spark
time