Chapter 6 Push Button Functions
103
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
Figure 6.23
Safety test
This function closes the output relays to perform the measurement(s) required
during the periodic safety check.
Activation of this function disables the
electrosurgical generator for normal operation. Refer to the maintenance
manual of the electrosurgical generator for a detailed safety test description.
NOTE
If the “Relays On button” will be pressed, the
electrosurgical generator remains in this mode until
the “Relays Off button” or “Cancel button” has
been pressed again. This enables a measurement
required by the periodic safety check.
1.
Select Menu screen
After pressing the “MENU push button”, press the “Next button” on the
“Select Menu screen” (see figure 6.24).
The “Select Menu screen” switches to the next page.
Press the “Safety Test button” on the “Select Menu screen”.
The screen changes from the “Select Menu screen” to the “Safety Test
screen” (see figure 6.25).
Software Version screen
Select Menu screen