Chapter 3 Installation and Connections
37
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
1.
Connecting the power cord
Confirm that the power of the electrosurgical generator is off. Connect the
power cord to the AC power socket
of the electrosurgical generator (see
figure 3.1).
2.
Connecting to the power outlet
Connect the power plug of the power cord directly to a grounded wall outlet
which meets the power requirements indicated on the electrical rating plate
on the rear panel of the electrosurgical generator (see figure 3.1).
Figure 3.1
3.5 Automatic mist & smoke evacuation
system/function (when using the compatible high
flow insufflation unit)
When the electrosurgical generator is combined with an Olympus high flow
insufflation unit (UHI-2/3) via a communication cable (MAJ-1871 or 1872), the
smoke and mist produced in the abdominal cavity can be evacuated
simultaneously with the electrosurgical output.
CAUTION
When using the automatic mist & smoke
evacuation system/function, also refer to the
instructions for use of the compatible high flow
insufflation unit (UHI-2/3).
Avoid
applying
excessive
force
to
the
communication
cable.
Otherwise,
wire
disconnection or other failure may result.
Do not use the 2-way cable (MAJ-1423) for
aeration and the extension cable for aeration in
combination with the electrosurgical generator.