Chapter 4 Inspection
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
4.5 Inspection of footswitch connection
1.
Connect a footswitch (e.g. double pedal footswitch) with the electrosurgical
generator on the rear panel according to chapter 3.6, “Connection of
footswitch”.
2.
Press the “FOOTSWITCH push button” on the right side of the touch-
screen as shown in figure 4.7.
The “Footswitch screen” appears.
Figure 4.7
3.
Press the “Double footswitch indicator” of the “MONOPOLAR 1 output
socket” (see figure 4.8).
The “Double footswitch button” changes to gray which indicates the
connection of the double pedal footswitch at the “MONOPOLAR 1 output
socket.”
4.
Disconnect the footswitch plug from the electrosurgical generator.
The “Double footswitch button” is black again.
Figure 4.8
Footswitch screen
Double footswitch button
Single footswitch button
FOOT
SWITCH