Chapter 6 Push Button Functions
105
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
Service
This function can be accessed by Olympus staff and service technician only. It
is password protected and not intended for customer use.
Output volume control
This function controls the volume of the output tone.
WARNING
Because the output tones play an important role of
noticing the output, do not lower the volume to an
inaudible level. If the output tone is inaudible,
output may not be detected by the user. This could
cause patient injury.
NOTE
The volume can be controlled either on the
touch-screen or using the volume control on the
rear panel of the electrosurgical generator.
The volume control on the touch-screen is not
available during output.
To control the volume during output, use the
volume control on the rear panel of the
electrosurgical generator.
The volume of the error tones (low, medium
and high priority errors) are not adjustable.
Refer to chapter 8, “Troubleshooting” for error
information and to “Alarm information” and
“Tone information” in the Appendix.
After pressing the “MENU push button”, press the “Plus button” or “Minus
button” in the “Volume” section of the “Select Menu screen” (see figure
6.26).