Chapter 2 Nomenclature and Functions
23
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
2.3 Rear panel
1.
Footswitch sockets
This socket connects the plug of a single or double pedal footswitch.
2.
Volume control
This knob is used for adjusting the output volume.
3.
Ventilation hole
Holes for air ventilation via a cooling fan; there are also ventilation holes on
each side of the electrosurgical generator.
4.
Equipotential bonding point
To increase electrical safety, this point is used for potential equalization. All
equipment housings that come into contact with the patient are electrically
connected in order to prevent low-frequency electrical currents from
endangering the patient in the event of a defect in the conventional
protective conductor system.
5.
AC power socket
This socket serves as a connection to the mains power supply via a power
cord.
6.
LINK-OUT socket
This socket connects the plug (14-pin) of a cable connected to peripheral
equipment.
7.
LINK-IN socket
This socket connects the plug (26-pin) of a cable connected to peripheral
equipment.
1.
2.
3.
4.
7.
6.
5.
1.