Chapter 8 Troubleshooting
119
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
What to do when an error code is displayed
If an error code is displayed, perform the indicated remedial actions below. If
the problem cannot be resolved by the described remedial action, contact
Olympus.
Error
no.
Error message
Possible cause
Remedial action
Electrodes of the HF
instrument may have no
proper tissue contact.
Ensure that the electrodes of the
HF instrument have proper
tissue contact.
Malfunction of the HF
instrument and / or the
connection cable.
Replace the HF instrument
and / or the connection cable.
E001
Open circuit
1. Check if the electrodes
of the instrument have
proper tissue contact.
2. If the problem persists,
replace the instrument.
3. If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
If the problem persists, contact
Olympus.
Electrodes of the HF
instrument may touch each
other.
Ensure that the electrodes of the
HF instrument do not touch each
other.
Malfunction of the HF
instrument and / or the
connection cable.
Replace the HF instrument
and / or the connection cable.
E002
Short circuit
1. Ensure the instrument
electrodes do not come into
contact with each other.
2. If the problem persists,
replace the instrument.
3. If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
If the problem persists, contact
Olympus.
Use of non-conductive fluid
during a bipolar cutting
procedure.
The active and / or neutral
electrode is within an air
environment.
Ensure that conductive fluid is
used during bipolar resection
procedure.
Always immerse the active
and / or neutral electrode within
the conductive fluid.
The bipolar Olympus HF
instrument has not been
properly connected to the
UNIVERSAL socket or
damaged connection cable.
Check the connection of the
bipolar Olympus HF instrument
to the connection cable and the
connection of the connection
cable to the UNIVERSAL socket
and / or replace the connection
cable.
The electrode might be
contaminated and
encrusted.
Check the electrodes for
contamination and encrustation
before use and, if necessary,
clean the electrodes.
E006
Non-conductive fluid
1. Ensure conductive fluid
is used for bipolar
resection.
2. If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
If the problem persists, contact
Olympus.