116
Chapter 8 Troubleshooting
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
Situation
Possible cause
Remedial action
The volume is set to an inaudible
level (e.g. due to high environmental
noise).
Increase the volume either on the
touch-screen within the “Menu screen”
or use the volume control on the rear
panel of the electrosurgical generator.
No sound is audible
during activation.
Malfunction of the electrosurgical
generator.
Stop using the electrosurgical
generator and press the power switch
to turn off the electrosurgical
generator. Contact Olympus.
The volume of the error-related
audible signal is not adjustable.
No action required.
The volume can not
be adjusted via the
volume control
within the “Menu
screen” or at the
rear panel.
Malfunction of the electrosurgical
generator.
Contact Olympus.
Improper connection of the
footswitch to one of the footswitch
sockets on the rear panel of the
electrosurgical generator or the HF
instrument to the output socket.
Check the footswitch and the HF
instrument for correct connection.
The footswitch or the handswitch of
the HF instrument and / or the
connection cable are damaged.
Check the footswitch or the handswitch
of the HF instrument and / or the
connection cable for damages and, if
necessary, replace the footswitch, the
HF instrument or the connection cable.
The incorrect footswitch pedal or
handswitch button is pressed.
Press the correct footswitch pedal or
handswitch button of the HF
instrument.
The electrosurgical generator is not
switched on.
Switch on the electrosurgical generator
with the power switch.
Another footswitch pedal or
handswitch button of the HF
instrument is pressed.
To activate the intended output,
release the current pressed footswitch
pedal or handswitch button of the HF
instrument.
The output is activated by the
peripheral equipment.
If the output of the peripheral
equipment is activated, the output of
the electrosurgical generator cannot be
activated simultaneously. Stop using
the peripheral equipment.
A window is displayed on the touch-
screen.
Press the “OK button” or “Cancel
button” to close the window or wait
until the window disappears
automatically after a few seconds.
The electrosurgical
generator does not
respond to
footswitch or
handswitch
activation.
The “All screen” or “Set screen is not
displayed on the touch-screen.
Return to the “All screen” or “Set
screen”.