Chapter 8 Troubleshooting
115
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
What to do when no error code is displayed
Perform the indicated remedial actions below. If the problem cannot be resolved
by the described remedial action, contact Olympus.
Situation
Possible cause
Remedial action
Improper connection of the power
cord to the AC power socket on the
rear panel of the electrosurgical
generator or to the grounded wall
outlet.
Check the power cord and the
grounded wall outlet for correct
connection.
The grounded wall outlet has wrong
or not output voltage.
Check the grounded wall outlet or use
an alternative grounded wall outlet.
The power cord is damaged.
Check the power cord for damages
and, if necessary, replace the power
cord.
The electrosurgical
generator does not
respond after
pressing the power
switch.
Malfunction of the electrosurgical
generator.
Contact Olympus.
The touch-screen
remains dark after
switching the
electrosurgical
generator on
(sound is audible
after switching on).
Malfunction of the touch-screen.
Contact Olympus.
An object is in contact with the touch-
screen.
Remove the object.
The touch-screen is not properly
calibrated.
Contact Olympus.
The touch-screen
cannot be
controlled.
Malfunction of the touch-screen.
Stop using the electrosurgical
generator and press the power switch
to turn off the electrosurgical
generator. Contact Olympus.
A push button is already pressed.
Release the push button.
The electrosurgical
generator does not
react when a push
button on the front
panel is pressed
during standby.
Malfunction of the electrosurgical
generator.
Contact Olympus.
The electrosurgical
generator does not
react when a (push)
button on the front
panel is pressed
during activation.
The (push) buttons are not available
during activation.
Release the footswitch or handswitch
to stop the activation.