198
Chapter 9 Function setup
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
Light sources
This operation sets the type of light source connected.
Other settings of the light source may cause improper color
and/or brightness in the image.
1.
Click “
” of the list box “Type”. The following light sources appear in the
pull-down menu.
2.
Click the setting value corresponding to the connected light source. The
selected setting value is displayed.
• When the CLV-180 is used, the setting is not necessary.
• When a light source other than CLV-180 is used, set up the
light source according to the table below. Please refer to the
instruction manual of the light source.
Setting value
Connected light source
CLV
CLV-160, CLV-U40
CLV-S
CLV-S40
CLE
Halogen light source
Table 9.4
Setting items
Settings
Brightness adjustment mode
AUTO
BRIGHTNESS
Center of the adjustable range
Table 9.5