Chapter 8 Installation and Connection
191
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
When a mobile workstation other than the WM-NP1 and
WM-WP1 is used or when no mobile workstation is used
1.
Confirm that the video system center is OFF.
2.
Connect the power cord provided with the video system center first to its AC
power inlet, then to the wall mains outlet.
3.
Connect the instruments listed in Table 8.25 to the wall mains outlet.
4.
Connect the instruments listed in Table 8.26 to the isolation transformer.
5.
Connect the power cord of the isolation transformer to the wall mains outlet.
Model
Product name
CLV-180, CLV-160, CLV-U40, CLV-S40,
Light source
OEV monitors
(OEV143, OEV203, OEV181H, OEV191, OEV191H)
Monitor
OEP printers
Video printer
Table 8.25 The devices to be connected directly to the wall mains outlet
Model
Product name
DVO-1000MD, DSR-20MD, SVO-9500MD
VCR
Other than OEP video printers
(UP-1800, UP-1850, UP-2900MD, UP-2950MD, UP-5200MD,
UP-5250MD, UP-21MD)
Video printer
Other than OEV video monitors
Monitor
Table 8.26 The devices to be connected to the isolation transformer