74
Chapter 28
| Appendix F: Fit Test Adapters |
28
Appendix F: Fit Test Adapters
a. The adapters must be kept clean and in good shape.
b. Disconnect and reconnect all tubes to make sure the connectors “snap” into place. Hold
the assembly by the tubes rather than the metal connector.
c. If the adapter had been dropped or crushed, a hairline crack might cause leakage.
d. To verify the adapter is air-tight, connect only the blue tube to the adapter with the
single hose connection. Submerge the adapter and hose connection completely under
water. Then press and hold the SEAL CHECK button for at least 5 seconds. Repeat this
several times. If bubbles are seen coming from the adapter, there is a leak in the
diaphragm or in the eight screws holding the top plate on. Replace the diaphragm and
tighten screws firmly. Conduct the test again to ensure no bubbles are seen.
e. Check all screws and connectors. If the screws on the top of the adapter are not tight,
or if the connector ports can be turned by hand, this means they are not tight enough
and may be introducing a leak into the system.