61
Chapter 21
| Service |
21
Service
o
Should service be needed for your instrument, contact the OHD Service Department. For
those instruments outside the U.S. contact OHD to find the nearest service center.
o
CAUTION: The QuantiFit2 is intended to be serviced only by authorized service personnel.
Troubleshooting and service procedures should only be performed by qualified
technicians.
o
If the QuantiFit2 appears to be tampered with, this may render the warranty null and void.
o
For service and operation questions, contact OHD: [email protected],
+1.205.980.0180.
Calibration and Maintenance
To ensure that the QuantiFit2 is working properly and within specification, a daily verification
must be performed before each day of testing, and an annual factory calibration must be done
once each year.
o
Calibration & Maintenance
·
The QuantiFit2 is an OSHA accepted Quantitative (QNFT) fit test instrument. It utilizes
the Controlled Negative Pressure (CNP) fit test method as outlined in the Respiratory
Protection Standard - 29 CFR 1910.134, Appendix A.
·
The QuantiFit2 is a highly accurate precision instrument that uses electronic pressure
sensors to detect minute changes to very low levels of negative pressure to an accuracy
of +/- 0.25%. To ensure the instrument operates at the parameters for which it is
designed, OHD requires the QuantiFit2 to receive annual maintenance and calibration
at an authorized service facility.
o
Reasons for Annual Maintenance & Calibration:
·
OSHA Mandate – 29 CFR 1910.134 Appendix A Part I.C.1.(b) – “The employer shall
ensure that QNFT equipment is kept clean, and is maintained and calibrated according
to the manufacturer’s instructions so as to operate at the parameters for which it was
designed.”
·
Best Practice – Annual calibration and maintenance is the accepted industry standard
for Industrial Hygiene and Occupational Health instrumentation.
·
Reliable Results – The QuantiFit2 is a precision instrument that is used to help ensure
employee safety. Critical decisions are made based on the results provided by the
QuantiFit2. All precision instruments require some degree of calibration to ensure the
measurements are accurate.
·
Limits Liability – Complying with the manufacturer’s guidelines establishes good faith
efforts towards employees’ health and safety. Failure to comply with established safety
protocols could be viewed as negligent.