Chapter 9
| Settings | Admin Menu
9
Settings
o
Challenge Pressure: The negative pressure rate at which a respirator is challenged
·
0.58 is used in the APR parameters
·
1.0 is used for CBRN and military function
·
1.5 is used for SCBA respirators for a
maximum challenge
o
Modeled Breathing Rate: The breath that is
emulated during the fit test
·
0.53 is used for APR settings
·
0.59 is used for CBRN and military function
·
0.93 is used for SCBA respirators
o
Passing fit factor: The minimum fit factor required in order to pass a test step and overall fit
factor.
Change Regulation
The QuantiFit2 is designed so that when a regulation is selected, all tests are automatically
executed with the correct settings. New regulations will be added in future updates as they
become available or are modified. For a more detailed explanation and description of the
regulations, please see
. The current regulations include:
o
OSHA
(Occupational Safety and Health Administration - US)
o
CSA
(Canadian Standards Association - Canada)
o
UK-HSE
(Health and Safety Executive - UK)
o
ISO
(International Standards Association)
System Reset
There are numerous reset options according to which information needs to be reset.
Clear People List
will remove all of the people in the current database. This would be helpful if
another entity will use the QuantiFit2, or to upload a fresh list of names for a new year of
testing.