30
Chapter 8
| Daily Verification |
8
Daily Verification
The daily verification is required for each day of
testing. The purpose is to confirm that the
instrument is working properly.
When turning on the instrument for the first time
each day, the Operator will first be prompted with
the daily verification message. Once the daily
verification has been performed for the day, this
message will no longer appear when the unit
cycles power.
The daily verification test measures the leak rate of the leak orifice on the triple tube assembly
at various pressure levels, and confirms that the diaphragm pump and other processes are
working correctly.
The daily verification may be performed at other times for other purposes. A Daily Verification
button can be found on the QuantiFit2 Home Screen.
o
Run a daily verification when the instrument moves to another location, especially with any
environmental changes.
o
Run a daily verification whenever the triple tube assembly is changed.
o
Troubleshooting:
·
If the instrument passes the daily verification, this confirms that the QuantiFit2 is
working properly.
·
If the instrument does not pass the daily verification, troubleshooting information will
be diplayed.
·
If the verification will not pass after troubleshooting steps, the unit may require service.
Contact OHD technical support (
).