Section 6
Fuse Replacement
•
See figures 39 and 40 for fuse location.
Unplug the power cord (floor models) or turn off main power before fuses are
replaced. Failure to do so could result in personal injury and/or property damage.
•
Use a flat blade screwdriver, turn the fuse holder counter-clockwise ¼ turn. Remove the
spring-loaded holder.
•
Replace the blown/broken fuses with Nuvo approved replacement fuses.
Use only Nuvo fuses if replacement is necessary. Failure to do so could result in
personal injury and/or property damage.
To reduce the risk of fire, replace the fuse only with the same type and electrical
rating as indicated on the attached table or on the fuse replacement label.
Note: Do not use light should it continue to blow fuses. Contact the factory
immediately
.
INPUT VOLTAGE
AMPERAGE RATING
P/N
VistOR PR, VistOR
PRO
(All mounting styles)
115/120 VAC
230/240 VAC
100 VAC
1.50 A, 250V
0.75A, 250 V
1.50 A, 250V
NV0001314
NV0001460
NV0001314
VistOR SP Spotlight
Standard
VistOR SPF Spot
Light
(All mounting styles)
115/120 VAC
230/240 VAC
100 VAC
115/120 VAC
230/240 VAC
2.00 A, 250V
1.00 A, 250 V
2.00 A, 250V
3.50 A, 250V
1.80 A, 250 V
NV0001396
NV0001461
NV0001396
NV0001347
NV0001462
VistOR EX & EXL
(All mounting styles)
115/120 VAC
230/240 VAC
100 VAC
1.50 A, 250V
0.75 A, 250 V
1.50 A, 250V
NV0001314
NV0001460
NV0001314