Section 3
Installation of Light Head to Arm
•
Insert the yoke/arm connector into the arm until the bolt-holes are aligned. Insert the ¼”-28 x
19
/
16
” bolt as shown being careful not to damage the wires, and secure with the ¼”-28 hex nut.
Tighten the nut securely, but do not over tighten.
•
Tighten the anti-rotation screw into the yoke/arm connector using the hex key provided.
•
One set of wires exits from the light head assembly, and two other sets exit from the arm/wire
junction assembly. Connect the wires with spade connectors, one from the arm and one from
the head as shown. Connect the wires with flag connectors on the switch box. Position the
connectors in the junction housing to allow room for the switch. Replace the switch assembly
cover and securely tighten the two screws. See figure 18.
•
Press the snap cover onto the top of the hex bolt as shown on figure 17.
FIGURE 18: Light Head to Arm Electrical Connections
Spade Connectors (2)
Flag Connectors (2)