Floor Stand Light Installation
•
•
Bolt the assembled legs onto the cast iron base using four ¼”-20 x 1 ¾” bolts and ¼” lock
washers. Position the floor base casting so that the set-screw in the base is at the rear of the
assembly. See figure 15.
•
Insert the upright pole fully into the floor base casting so that the anti-rotation slot sits over the
pin in the casting. Rotate the pole until the slots slides fully over the pin, lowering the pole. The
power cord (not shown on figure) should be in line with the set-screw. Once the pole is properly
positioned, securely tighten the set-screw using the 1/8” hex key provided.
Do not install the extension arm with the light head attached. Installation with the light
head attached can cause damage to the light. Refer to the procedure for installing the head
to the arm after the floor mount is assembled and the arm is in place.
FIGURE 15: Floor Stand Light Installation
Anti
-rotation
Slot
Base Casting
P/N 1001170
Power Cord
Exit location
Set Screw
P/N 0001128
Caster (2)
Brake
Anti-rotation
Pin
P/N 0001208
Leg (2)
P/N 1001208
End Cap (4)
P/N 0001206
Bolt (4)
P/N 0001309
Flange Nut (4)
P/N 0001074
Caster (2)
P/N 0001073
Upright Assembly
P/N 1000125
Adjustment
Nut (4)
Lock Washer (4)
P/N 0001122
P/N 0001363