P
age 69 of 85
5368 KUHL ROAD
●
ERIE, PENNSYLVANIA 16510
●
PHONE 814-899-4220
●
FAX 814-899-1410
A member of the Medical Illumination International Group
Section 8
Section 8: Exploded Views and Parts List
VistOR PR/PRO and VistOR EX/EXL Electric Schematic
12 VAC
Neutral
12 VAC
Lamp
Type T4, GY 6.35 Base
75W, 12VAC (VistOR EX and VistOR EXL)
100W max 12VAC (VistOR PR and VistOR PRO)
Fuse: ¼”x1 ¼”
1.50 A, 250 V
Slow Blow
(.75A, 250 V
Slow Blow)
Switch, toggle
SPST, On-Off Circuit
20A, 125 VAC
10A, 250 VAC
FIGURE 44:
Electric Schematic for VistOR PR/PRO and VistOR EX, 120, 220/230, & 240 VAC
Fuse: ¼”x1 ¼”
1.50 A, 250 V
Slow Blow
(.75A, 250 V
Slow Blow)
220/230 VAC
120 VAC, 50/60 Hz
220/230 VAC 50/60 Hz
240 VAC, 50/60 Hz
Power Transformer
Split bobbin type
Rated @ 100 VA, 12VAC
240 VAC
Содержание VISTOR
Страница 1: ...VistOR Installation Service Manual...