Page 10 of 85
5368 KUHL ROAD
●
ERIE, PENNSYLVANIA 16510
●
PHONE 814-899-4220
●
FAX 814-899-1410
A member of the Medical Illumination International Group
Section 2
Specifications
VistOR PR Specifications
Mechanical
Parameter
Value
Weight
Head
Arm
Arm (Floor)
Wall Bracket Assy.
Floor Stand Assy.
Single Ceiling Mount
Dual Ceiling Mount
Approximately 9.10 lbs. (4.14 kg).
Approximately 12.5 lbs. (5.68 kg).
Approximately 11.4 lbs. (5.18 kg).
Approximately 3.00 lbs. (1.36 kg).
Approximately 43.0 lbs. (19.50 kg).
Approximately 14.5 lbs. (6.58 kg).
Approximately 16.0 lbs. (7.26 kg).
Dimensions
Head
Arm
Arm (Floor)
Floor Stand
15.8” (40.13 cm) Dia. X 6.0 (15.24 cm) High
57.0” (144.8 cm) Long
38.5” (97.79 cm) Long
74.5” (189.23 cm) High
Rotations
Articulating Arm Vertical Movement
Articulating Arm Horizontal Movement
Articulating Arm/Yoke Interface
Yoke/Light Head Interface
+/- 40 Degrees travel
Approximately 345 Degrees
Approximately 540 Degrees
Approximately 190 Degrees
Electrical
Parameter
Value
Voltages
Input Voltage
Bulb Voltage
110 - 125 VAC 50/60 Hz
210 - 230 VAC 50/60 Hz
11 – 12 VAC 90 W
Bulb life
2,000 hours (average)
Optical
Parameter
Value
Reflector
15.0“ (380mm) Dia., Polished, Alzak
Performance
Central Illuminance
Color Temperature
Field Size
Irradiance
CRI
Approximately 4,000 ft-candles (43000 Lux) @ 39.3” (1m)
(4,300 ft-candles (46,285 Lux) @ 36” (91.44 cm) )
Approximately 4,000 °K
6.00” (152.4 mm) Diameter @ 39” (914.4 mm).
155.2 W/m
2
90
Environmental
Parameter
Value
Operating Temperature
Storage Temperature Range
41 -104 ° F (5-40 °Celsius)
42 -113 ° F (5-45 °Celsius)
Humidity
10 - 90% Relative Humidity
Содержание VISTOR