Section 6
Initial Lamp Installation/Lamp Replacement
VistOR EX & EXL
The VistOR lights operate at high temperatures. Allow the unit or units to cool at
least 30 minutes before performing any maintenance. Failure to do so could result in
personal injury
.
Installation and repair of this equipment should be performed by qualified persons
only. Nuvo does not warranty any damage or injuries occurring as a result of improper
installation.
Unplug the power cord (floor models) before lamps are replaced. Failure to do so
could result in personal injury and/or property damage.
•
Twist the rear cover counter clockwise approximately 7 degrees to unlock the rear cover and
remove it from the head as shown in figure 37.
•
Remove the lamp from its package. Use a tissue or glove when handling the lamp to avoid
touching the glass with your fingers. Oil from your hands will shorten the life of the lamp. If
contact is made, clean the lamp with alcohol after installing it into the holder.
•
Grasp the lamp by its base and carefully insert the metal pins into the ceramic socket at the
end of the lamp holder. Make sure the lamp is fully seated in the lamp holder (pins will
remain partially exposed).
•
Insert the two metal jacks on the lamp holder into the mating plugs located in the light head.
Make sure the lamp holder is fully and evenly seated. Replace the rear housing cover.
•
When replacing a lamp, use Nuvo P/N NV0001257 for the VistOR EX & EXL
Diagnostic/Specialty light, and use light bulb P/N NV0001717 for the VistOR O.B.
Do not touch the lamp directly. Body oils may significantly lower the life expectancy
of the lamp and cause equipment damage. If contact is made, clean the lamp with
alcohol after installing it into the holder.