Page 18 of 85
5368 KUHL ROAD
●
ERIE, PENNSYLVANIA 16510
●
PHONE 814-899-4220
●
FAX 814-899-1410
A member of the Medical Illumination International Group
Section 3
Ceiling Rod Calculation
Dual Ceiling Mount
Use the following table to select the correct length ceiling rod for your application.
Ceiling Rod Calculation, Dual Ceiling Mount
Ceiling Mounting
Height
Ceiling Rod
Length
“X” Value
“Y” Value
Head room to bottom
of extension arm
(y value-x value)
7’6”- 8’6”
(2,29- 2,59 m)
16”(406,4 mm)
22.5”(571,5 mm)
89.5-102.5”
(2,27-2,57 m)
67.5-79.5”
(1,70-2,00 m)
8’7”- 9’6”
(2,62 -2,89 m)
25”(635,0 mm)
31.5”(800,1 mm)
102.5-114.5”
(2,60-2.87 m)
71.5-83.5”
(1,80-2,11 m)
9’7”- 10’6”
(2,92- 3,20 m)
35”(889,0 mm)
41.5”(1054,1 mm)
114.5-126.5”
(2,90-3.18 m)
73.5-85.5”
(1,85-2,16 m)
Note: The VistOR PR / VistOR SP Combination Light is shown for reference.
The dimensions
indicated also apply to all other combination of fixtures and dual mounting of the same light.
Dimensions are approximate
*Length measured when Arm is in horizontal position.
NOTE: Longer ceiling
rods are available for
higher ceilings. See
page 23 for extended
ceiling mount
installation instructions.
FIGURE 2: Dual Ceiling
Mount Ceiling Rod Calculation
“X” VALUE
(TO BOTTOM OF
EXTENSION ARM)
“Y” VALUE
(DISTANCE TO FLOOR)