P
age 66 of 85
5368 KUHL ROAD
●
ERIE, PENNSYLVANIA 16510
●
PHONE 814-899-4220
●
FAX 814-899-1410
A member of the Medical Illumination International Group
Section 6
Maintenance Schedule
Function
Procedure
Light bulb/Lamp
Holder
Ensure that light bulb is seated properly in socket. Examine socket and
wiring for signs of heat degradation. Inspect wiring for signs of chafing.
Replace any damaged parts.
Glass lens/Reflector
Check glass lens and reflector to assure there are no chips, cracks, or
other damage. Do not use equipment if glass parts are damaged.
Replace damaged parts immediately.
Bolts and nuts
Check to see that all mounting and attachment screws, washers, etc. are
in place and securely tightened. Replace any missing screws and re-
tighten as required.
Casters
Ensure that casters are seated properly on the base assembly. Examine
the base for any damage.
Moving
joints/Adjustments
Check to make sure all moving joints function properly along the
mounting system and head and arm system. If the articulating arm does
not position properly (drifts from original position) refer to arm adjustment
on figure 41, page 66. If this does not solve the problem contact
customer service as the unit may require factory repair.
Overall appearance
Check the general aesthetics. Units should be kept clean and dust free.
Clean and dust as necessary.
Note: Maintenance schedules vary for each light depending on usage and operating
instructions. An annual inspection of the equipment is recommended at a minimum.
Note: Nuvo, Inc. recommends that the maintenance records for this equipment be
kept on file at the health care facility.