![Nustep TRS 4000 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/nustep/trs-4000/trs-4000_service-manual_1718276061.webp)
TRS 4000
[email protected] - (800) 322-4434 - (734) 769-3939 x 5
[email protected] - (800) 322-4434
EC Plunger Plate & EC Magnet Slide Bracket
NOTE: These parts come off sequentially. Once you get to the part you need to replace,
replace that part and skip to the other steps as the instruction tells you.
Tools Required
Channel lock pliers
3/4” deep socket and new locking ring
3/16” hex key
Hammer
SIlicone/grease (white or clear)
Safety glasses
2”x4” wood
1. Remove covers.
See
Cover Removal
instructions.
2. Remove poly-v belt or poly-v J6 flex belt.
For S/N < 428325, remove poly-v belt.
See
Poly-V Belt
instructions.
For S/N > 428326, remove poly-v J6 flex belt.
See
Poly-V J6 Flex Belt
instructions.
3. Remove load lever.
See
Load Lever
instructions.
4. Remove self-locking ring. Ring
will be destroyed during this step.
5. Pull plunger plate out of frame.
NOTE: if you are just replacing plunger
GOTO Step 9.
42033 EC Plunger & Slide Brkt 5-01-05
53
To avoid injury, wear safety glasses.
!
CAUTION