INTERFACE A (DB-9)
PAGE
|
87
Temperature OK – High (+5V) PWR LED – Flashing
Over-temp Fault – Low (0V) DC Voltage Fault – High (+5V)
Possible Causes:
•
A PWM Sense fault or control board failure has occurred.
If this fault occurs, the laser requires service—contact NOVANTA Customer Service or a NOVANTA
Authorized Distributor.
Symptom:
•
The power supply is connected, and DC voltage is correct, but the following LED and I/O states
exist:
LASE LED – Off
Lase Indicator – Low (0V)
Laser OK – Low (0V)
Voltage OK – Low (0V) RDY LED – Off
Laser Ready – Low (0V)
Temperature OK – High (+5V) PWR LED – off
Over-temp Fault – Low (0V) DC Voltage Fault – High (+5V)
Possible Causes:
•
The 20 A fast-acting fuse has blown.
To replace the fuse, open the inline fuse holder (located on the positive DC power cable) by twisting the
upper half 1/4 turn counterclockwise. Remove the fuse and replace it with a 20 A, 250 V fast-acting fuse
(Cooper Bussmann ABC-20-R, Littelfuse 314020, or equivalent).
Interface A (DB-9)
Symptom:
•
The built-in +5 VDC, 250 mA voltage output on Pin 5, DC Out, was accidentally shorted to
ground.
Possible Causes:
•
If Pin 5, DC Out, is shorted to ground, the OEM v30 laser will not be damaged as the voltage
regulator is protected by a current-limiting circuit. Remove the short circuit to re-enable the
voltage output.
Interface B (RJ45)
Symptom:
Содержание v30
Страница 1: ...ENGINEERED BY SYNRAD v30 Laser User Manual...
Страница 16: ...OEM V30 LASER LABEL LOCATIONS PAGE 16 OEM v30 Laser label locations...
Страница 22: ...DECLARATION OF CONFORMITY PAGE 22 Declaration of conformity Figure 2 Declaration Document...
Страница 64: ...INTERFACE CONNECTIONS PAGE 64 Figure 3 16 Laser OK output to PLC input...
Страница 90: ...BEAM DELIVERY OPTICS PAGE 90 This page is left blank intentionally...
Страница 100: ...INDEX PAGE 1 This page is intentionally left blank...