Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
62
If any time-change component is not replaced, original
performance may not be demonstrated.
When attaching the probe or replacing a component, take care
not to damage the probe connector.
The name and criterion of replacement of each component of the probe are as follows:
No.
Name
Recommended Replacement Period and Criterion
1
Probe Filter
1 month or 500 operation hours
2
Probe Tube
3 months or 1000 operation hours
3
Probe Connector
6 months
*
The recommended replacement period is not the guaranteed lifetime.
Replacement of Probe Filter
(1)
Turn the connection to the probe connector and remove the probe filter.
(2)
Attach a new probe filter.