Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
63
Replacement of Probe Tube and Probe Connector
(1)
Remove the probe filter.
(2)
Remove the probe tube and probe connector.
*
To remove the probe tube and probe connector, pull them horizontally. If you try to
remove them obliquely, the probe connector may break.
*
If it is difficult to remove them, tear the probe tube little by little using a cutter, and
remove them. In that case, take care not to get hurt or damage the probe connector.
(3)
Connect a new probe tube and a new probe connector.
*
If it is difficult to insert the probe connector into the probe tube, expand it using a
tool or the like and then pull in the probe connector.
*
If the probe tube is expanded too much using a tool, the probe connector will be apt
to slip out. Carry out the work while checking the effect.