Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
45
(2)
Enter an atmospheric pressure in the measurement environment and press “Enter” for
determination.
Button
Description
Cancel
Cancellation of alteration
Clear
Deletion of all characters
Enter
Determination
BS
Deletion of one character
*
From a difference from the set pressure, the pressure of measurement gas is calculated
and a correction is made. By setting “Ambient Press.” according to the altitude etc. of
the measurement environment, a more accurate normalization correction can be made.
Saving Interval
Measurement results are saved in the built-in micro SD card. Set an interval for the saving.
(Default Setting: 1 sec.)
(1)
Press the numerical button.
(2)
Enter a saving interval and press “Enter” for determination.
Presettable Range
: 1 - 6000 sec.
Button
Description
Cancel
Cancellation of alteration
Clear
Deletion of all characters
Enter
Determination
BS
Deletion of one character